Overview

The table below shows results from a compound interference study. We tested a panel of 24 compounds in a QC assay using our AlphaLISA™ toolbox bead products. The information below can be used for guidance as to whether Tween-20 might interfere in assays involving particular bead products, and at what Tween concentration interference might occur. These data are derived from single experiments and should serve more as a guide rather than a precise value, as it is unlikely you will be using the exact same assay design as used to generate the data below. Interference concentrations may vary depending on the assay components. It is possible your assay will tolerate higher or lower concentration than what is shown.

    Tested up to 3%
Catalog Number Product No effect (10%) 50% loss % of inhibition at max conc

AL125
Streptavidin AlphaLISA Acceptor Beads >3 >3 0
AL126
Protein L AlphaLISA Acceptor Beads
>3 >3 0
AL127 Anti-FITC AlphaLISA Acceptor Beads >3 >3 0
AL128 (biotin) Anti-6xHis AlphaLISA Acceptor Beads >3 >3 0
AL128 Anti-6XHis AlphaLISA Acceptor Beads >3 >3 0
AL129
Anti-V5 AlphaLISA Acceptor Beads
>3 >3 0
AL130 Anti-mouse IgM AlphaLISA Acceptor Beads >3 >3 0
AL131
Anti-chicken IgY AlphaLISA Acceptor Beads
>3 >3 0
AL132 Anti-sheep IgG AlphaLISA Acceptor Beads >3 >3 0
AL133 Anti-GFP AlphaLISA Acceptor Beads >3 >3 0
AL134
Anti-MBP AlphaLISA Acceptor Beads
>3 >3 0
AL136
Strep-Tactin AlphaLISA Acceptor Beads
>3 >3 0
AS102 Protein A Alpha Donor Beads >3 >3 0
AS103 Anti-FLAG Alpha Donor Beads >3 >3 0
AS104 Anti-mouse IgG Alpha Donor Beads >3 >3 0
AS105 Anti-rabbit IgG Alpha Donor Beads >3 >3 0
AS106 Strep-Tactin Alpha Donor Beads >3 >3 0
AL900
AlphaLISA TruHits kit 2 ug/mL
>3 >3 0
AL900 AlphaLISA TruHits kit 20 ug/mL >3 >3 0


Refer to the figure below for more detail as to what the values in the table above indicate. Please note that the figure is provided as a general example to explain how the values in our interference tables were determined. The table above indicates that an assay testing the effect of Tween-20 using anti-6X His AlphaLISA Acceptor beads (with glutathione Donor beads) showed little effect on the assay signal at final Tween-20 concentrations up to 3%.

 

tween-20-and-bead-interference

Figure: Key for how the values in our interference tables are derived. The “no effect” concentration was obtained by extrapolating the MAX - 10% MAX counts. Refer to the table above for actual interference values for each toolbox bead product.