The table below shows results from a compound interference study. We tested a panel of 24 compounds in a QC assay using our AlphaLISA™ toolbox bead products. The information below can be used for guidance as to whether EDTA might interfere in assays involving particular bead products, and at what EDTA concentration interference might occur. These data are derived from single experiments and should serve more as a guide rather than a precise value, as it is unlikely you will be using the exact same assay design as used to generate the data below. Interference concentrations may vary depending on the assay components. It is possible your assay will tolerate higher or lower concentration than what is shown.

 

 

 

Tested up to 0.1M 

Catalog Number 

Product 

No effect (10%) 

50% loss 

% of inhibition at max conc 

AL125 

Streptavidin AlphaLISA Acceptor Beads 

4.3E-02 

>0.1 

38 

AL126 

Protein L AlphaLISA Acceptor Beads 

4.0E-02 

>0.1 

19 

AL127 

Anti-FITC AlphaLISA Acceptor Beads 

1.0E-03 

1.2E-02 

88 

AL128 (biotin) 

Anti-6xHis AlphaLISA Acceptor Beads 

2.0E-03 

4.6E-03 

97 

AL128 (GSH) 

Anti-6XHis AlphaLISA Acceptor Beads 

1.2E-03 

3.0E-03 

97 

AL129 

Anti-V5 AlphaLISA Acceptor Beads 

2.2E-02 

>0.1 

49 

AL130 

Anti-mouse IgM AlphaLISA Acceptor Beads 

1.9E-03 

9.7E-02 

50 

AL131 

Anti-chicken IgY AlphaLISA Acceptor Beads 

1.8E-02 

>0.1 

32 

AL132 

Anti-sheep IgG AlphaLISA Acceptor Beads 

2.1E-02 

>0.1 

28 

AL133 

Anti-GFP AlphaLISA Acceptor Beads 

1.7E-03 

1.9E-02 

62 

AL134 

Anti-MBP AlphaLISA Acceptor Beads 

3.0E-02 

3.7E-02 

58 

AL136 

Strep-Tactin AlphaLISA Acceptor Beads 

>0.1 

>0.1 

AS102 

Protein A Alpha Donor Beads 

1.1E-02 

>0.1 

40 

AS103 

Anti-FLAG Alpha Donor Beads 

6.9E-03 

3.0E-02 

81 

AS104 

Anti-mouse IgG Alpha Donor Beads 

1.5E-03 

6.0E-02 

63 

AS105 

Anti-rabbit IgG Alpha Donor Beads 

4.3E-02 

>0.1 

38 

AS106 

Strep-Tactin Alpha Donor Beads 

>0.1 

>0.1 

AL900 

AlphaLISA TruHits kit 2 ug/mL 

8.0E-03 

>0.1 

34 

AL900 

AlphaLISA TruHits kit 20 ug/mL 

2.0E-02 

>0.1 

21 

 

Refer to the figure below for more detail as to what the values in the table above indicate. Please note that the figure is provided as a general example to explain how the values in our interference tables were determined. The table above indicates that an assay testing the effect of EDTA using anti-6X His AlphaLISA Acceptor beads (with glutathione Donor beads) showed little effect on the assay signal at final EDTA concentrations up to 1.2 mM. EDTA did show a half-maximal inhibition of signal when used at a final concentration of 3 mM. When 0.1 M EDTA was included in the assay, 97% signal inhibition was observed.

 

edta-bead-fig1.png

Figure: Key for how the values in our interference tables are derived. The “no effect” concentration was obtained by extrapolating the MAX - 10% MAX counts. Refer to the table above for actual interference values for each toolbox bead product.

For research use only. Not for use in diagnostic procedures.

The information provided above is solely for informational and research purposes only. Revvity assumes no liability or responsibility for any injuries, losses, or damages resulting from the use or misuse of the provided information, and Revvity assumes no liability for any outcomes resulting from the use or misuse of any recommendations. The information is provided on an "as is" basis without warranties of any kind. Users are responsible for determining the suitability of any recommendations for the user’s particular research. Any recommendations provided by Revvity should not be considered a substitute for a user’s own professional judgment.