LANCE Instrument Options and Settings

Overview

LANCE™ and LANCE™ Ultra assays require a microplate reader that is capable of time-resolved fluorescence. We recommend filter-based fluorometers, though monochromator instruments can be used. Use of monochromator instruments may result in decreased assay sensitivity. We recommend using Revvity's VICTOR™ X, ViewLux™ (for high throughput), or EnVision™ Multilabel Plate Reader (view our instrument product pages). Please check that you have the correct filters for this assay. If you are using an instrument manufactured by another company, we recommend that you contact their technical support team to see if they have optimized settings, filter recommendations, etc. for a LANCE TR-FRET assay.

Use an excitation wavelength of 320 or 340 nm to excite the LANCE Europium chelate. We recommend you read this assay in dual emission mode, detecting both the emission from the Europium donor fluorophore at 615 nm, and the acceptor fluorophore (at 665 nm for APC, ULight™ dye, and Alexa Fluor 647). The raw FRET signal at 665 nm can be used to process your data. The ability to analyze the signal at 615 nm can be helpful when troubleshooting.

Recommended Instrument Settings for the LANCE cAMP Assay

Parameter VICTOR EnVision ViewLux EnSpire™ TRF option
Flash Energy Area High n/a n/a  
Flash Energy Level 150 100% 600,000  
Excitation Filter 320/340 UV2 320 DUG11 (UMB,AMC) UV2 320
Integrator Cap 3 n/a n/a  
Integrator Level 2x LANCE High Count 615 n/a n/a  
Emission Filter 1) 615 nm 1) 203 - Eu 615 1) 618/8 (Eu) 1) 615 nm**

Emission Filter
2) 665 nm 2) 205 - APC 665 2) 671/8 (LANCE) 2) 665 nm** 
Delay Time 50 µs 60 µs 50 µs 60 µs
Readout speed, gain and binning n/a n/a Medium, High, and 2x  
Number of Flashes n/a 100 n/a  
Window 100 µs (200 µs) 100 µs (200 µs) 354 µs 200 µs
Mirror module n/a 462 (D400/D630) or 412 (D400) n/a  
Cycle 2000 µs 2000 µs n/a  

 

Recommended Instrument Settings for the LANCE Ultra Kinase Assay

Parameter VICTOR EnVision ViewLux* EnSpire™ TRF option
Flash Energy Area High n/a n/a  
Flash Energy Level 150 100% 800,000  
Excitation Filter 320/340 UV 320/340 DUG11 (UMB, AMC) UV2 320
Integrator Cap 2 (or 3**) n/a n/a  
Integrator Level 2x the setting in LANCE High Count 615 label n/a n/a  
Emission Filter 1) 615 nm 1) 203 - Eu 615 1) 618/8 (Eu) 1) 615 nm**
Emission Filter 2) 665 nm 2) 205 - APC 665 2) 671/8 (LANCE) 2) 665 nm**
Delay Time 50 µs 90 µs 50 µs 60 µs
Readout speed, gain and binning n/a n/a medium, high, and 2x  
Measurement Time n/a 100 (200***) flashes 20 s exposure time  
Window 100 µs (200-300 µs***) 100 µs (200-300 µs***) 354 µs 200 µs
Mirror module n/a 402/412 (D400) or 452/462/662 (D400/D630) Mirror 2 (UV dichroic)  
Cycle 1000 µs 2000 µs n/a  


* ViewLux with flat field correction, bias correction, cosmic ray detection, excitation energy compensation

** The EnSpire Multi-Label plate reader uses monochromator-based emission. The wavelengths given represent the monochromator settings. The measurement is carried out first with one monochromator setting for the whole plate, then with the second monochromator setting in a subsequent measurement step.

*** If signal too low with 2 or 100

  • Laser settings for LANCE assays on an EnVision 
    Delay: 50 µs (you could measure delay 10µs to get slightly improved S/B) 
    Window time: 100 µs (it is important not to use too long window time) 
    Time between flashes: 16600 µs 
    Focus height: 6.5 mm