Skip to main content
Menu
US

Sodium fluoride and bead interference

Overview

The table below shows results from a compound interference study. We tested a panel of 24 compounds in a QC assay using our AlphaLISA™ toolbox bead products. The information below can be used for guidance as to whether sodium fluoride might interfere in assays involving particular bead products, and at what sodium fluoride concentration interference might occur. These data are derived from single experiments and should serve more as a guide rather than a precise value, as it is unlikely you will be using the exact same assay design as used to generate the data below. Interference concentrations may vary depending on the assay components. It is possible your assay will tolerate higher or lower concentration than what is shown.

    Tested up to 0.01M
Catalog Number Product No effect (10%) 50% loss % of inhibition at max conc
AL125 Streptavidin AlphaLISA Acceptor Beads >0.01 >0.01 0
AL126 Protein L AlphaLISA Acceptor Beads >0.01 >0.01 0
AL127 Anti-FITC AlphaLISA Acceptor Beads >0.01 >0.01 0
AL128 (biotin) Anti-6xHis AlphaLISA Acceptor Beads >0.01 >0.01 0
AL128 (GSH) Anti-6XHis AlphaLISA Acceptor Beads >0.01 >0.01 0
AL129 Anti-V5 AlphaLISA Acceptor Beads >0.01 >0.01 0
AL130 Anti-mouse IgM AlphaLISA Acceptor Beads >0.01 >0.01 0
AL131 Anti-chicken IgY AlphaLISA Acceptor Beads >0.01 >0.01 0
AL132 Anti-sheep IgG AlphaLISA Acceptor Beads >0.01 >0.01 0
AL133 Anti-GFP AlphaLISA Acceptor Beads >0.01 >0.01 0
AL134 Anti-MBP AlphaLISA Acceptor Beads >0.01 >0.01 0
AL136 Strep-Tactin AlphaLISA Acceptor Beads >0.01 >0.01 0
AS102 Protein A Alpha Donor Beads >0.01 >0.01 0
AS103 Anti-FLAG Alpha Donor Beads >0.01 >0.01 0
AS104 Anti-mouse IgG Alpha Donor Beads >0.01 >0.01 0
AS105 Anti-rabbit IgG Alpha Donor Beads >0.01 >0.01 0
AS106 Strep-Tactin Alpha Donor Beads >0.01 >0.01 0
AL900 AlphaLISA TruHits kit 2 ug/mL >0.01 >0.01 0
AL900 AlphaLISA TruHits kit 20 ug/mL >0.01 >0.01 0

 

Refer to the figure below for more detail as to what the values in the table above indicate. Please note that the figure is provided as a general example to explain how the values in our interference tables were determined. The table above indicates that an assay testing the effect of sodium fluoride using anti-6X His AlphaLISA Acceptor beads (with glutathione Donor beads) showed little effect on the assay signal at final sodium fluoride concentrations up to 0.01 M.

sodium-fluoride-bead-interference

Figure: Key for how the values in our interference tables are derived. The “no effect” concentration was obtained by extrapolating the MAX - 10% MAX counts. Refer to the table above for actual interference values for each toolbox bead product.


 

For research use only. Not for use in diagnostic procedures.

The information provided above is solely for informational and research purposes only. Revvity assumes no liability or responsibility for any injuries, losses, or damages resulting from the use or misuse of the provided information, and Revvity assumes no liability for any outcomes resulting from the use or misuse of any recommendations. The information is provided on an "as is" basis without warranties of any kind. Users are responsible for determining the suitability of any recommendations for the user’s particular research. Any recommendations provided by Revvity should not be considered a substitute for a user’s own professional judgment.