The AAV9 Capsid kit is designed for the quantitative measurement of Adeno-associated virus serotype 9 (AAV9) particles in both cell lysates and cell supernatants.
For research use only. Not for use in diagnostic procedures. All products to be used in accordance with applicable laws and regulations including without limitation, consumption and disposal requirements under European REACH regulations (EC 1907/2006).
Adeno-associated virus (AA) vectors are the leading platform for gene delivery for the treatment of a variety of human diseases. AAV Serotype 9 (AAV9) when used as a vector in gene therapy has been particularly developed for the treatment of neurological diseases mainly because of its ability to cross the blood-brain barrier to target the central nervous system (astrocytes and neurons). It also exhibits a tropism for other tissues such as liver, muscle or lung. The AAV9 kit is designed to detect and quantify AAV9 particles in an easy-to-use, no-wash format. The simple and robust procedure benefits from increased throughput compared to ELISA.
Application |
Bioprocessing
|
---|---|
Brand |
HTRF
|
Detection Modality |
HTRF
|
Product Group |
Kit
|
Sample Volume |
5 µL
|
Shipping Conditions |
Shipped in Dry Ice
|
Target Class |
Viral Particles
|
Technology |
TR-FRET
|
Unit Size |
500 Assay Points
|
The Adeno-Associated Virus serotype 9 (AAV9) assay measures AAV9 capsid in cell supernatant or cell lysate. The assay uses two labeled antibodies: one coupled to a donor fluorophore, the other to an acceptor. In presence of AAV9 capsid in a cell extract or supernatant the addition of these conjugates brings the donor fluorophore into close proximity with the acceptor, and thereby generates a FRET signal. Its intensity is directly proportional to the concentration of the capsid present in the sample and provides a means of assessing any changes caused by experimental variability in a no-wash assay format.
The AAV9 capsid assay protocol, using a 384-well small volume white plate, is described on the right. 5 µL of sample or standard and 5 µL of diluent are dispensed directly into the detection plate for detection by HTRF reagents. The antibodies labeled with HTRF fluorophores were added before an overnight incubation. The assay can be run in up to a 1536-well format by simply resizing each addition volume proportionally.
Sample size | 5 µL |
---|---|
Final assay volume | 20 µL |
Time to results | Overnight at RT |
Detection limit (LOD) in diluent | 1.78E+08 VP/mL |
Dynamic Range | 1.06E+09 – 2.00E+11 VP/mL |
Sample compatibility |
From raw harvest material to the final product Supernatant, Cell Lysate (LB#3) |
Sample |
Mean [AAV9] (VP/mL) |
CV |
---|---|---|
1 |
1.60E+11 |
3% |
2 |
4.00E+10 |
3% |
3 |
1.00E+10 |
3% |
|
Mean CV |
3% |
Each of the 3 samples was measured 24 times, and % CV was calculated for each sample.
Sample |
Mean [AAV9] (VP/mL) |
CV |
---|---|---|
1 |
1.60E+11 |
3% |
2 |
4.00E+10 |
8% |
3 |
1.00E+10 |
9% |
|
Mean CV |
7% |
Each of the samples was measured in 3 independent experiments performed by different operators, and % CV was calculated for each sample.
The excellent % of recovery obtained from these experiments shows the good dilution linearity of the assay. Samples were AAV9 empty capsids serially diluted with the kit DIL5 diluent.
Dilution factor |
[AAV9] Expected (VP/mL) |
[AAV9] Mesured (VP/mL) |
Dilution Recovery |
---|---|---|---|
Neat |
- |
1.63E+11 |
100% |
2 |
8.14E+10 |
8.11E+10 |
100% |
4 |
4.07E+10 |
3.64E+10 |
89% |
8 |
2.03E+10 |
1.84E+10 |
91% |
16 |
1.02E+10 |
9.39E+09 |
92% |
Mean |
|
|
94% |
Different cell lines (SF9 and HEK293), representative of AAV manufacturing processes, were seeded in T175 flasks in complete culture medium at 37°C, 5% CO2. The cells were then lysed with 3 mL of lysis buffer #3 (1X) for 30 minutes at RT under gentle shaking.
AAV9 capsid at 5.0E+10 VP/mL was spiked in several dilutions of SF9 or HEK293 cell lysates not expressing AAV capsids (corresponding to different amount of total protein concentration, mg/mL). All samples, including lysates without capsid added were measured in the assay.
As expected, no AAV9 was detected in SF9 or HEK293 control cell lysates. However, for an accurate quantification of AAV9 concentration it is important to ensure that total protein concentration is below 1 mg/mL for SF9 cell lysates and HEK293 cell lysates by using dilution with lysis buffer #3.
At higher cell lysate total protein concentration, there is an interference that could affect the accuracy of the AAV9 quantification.
[AAV9] (VP/mL) |
[Total Protein] SF9 cell lysates Spiked Sample (mg/mL) |
Recovery |
---|---|---|
5.0E+10 |
1 |
102% |
0.5 |
96% |
|
0.25 |
99% |
|
0.125 |
99% |
|
0.05 |
92% |
[AAV9] (VP/mL) |
[Total Protein] HEK293 cell lysates Spiked Sample (mg/mL) |
Recovery |
---|---|---|
5.0E+10 |
1 |
97% |
0.5 |
98% |
|
0.25 |
97% |
|
0.125 |
97% |
|
0.05 |
98% |
Three different concentrations of AAV9 capsid were independently mixed with SF9 or HEK293 cell lysates diluted in lysis buffer #3 at 1 mg/ml. The measured concentrations were compared to the expected values and expressed as percentage of recovery.
Sample |
[AAV9] Standard (VP/mL) |
SF9 cell lysate (1 mg/ml) |
---|---|---|
Recovery |
||
1 |
1.16E+11 |
99% |
2 |
5.61E+10 |
100% |
3 |
2.64E+10 |
93% |
|
Mean CV |
97% |
Sample |
[AAV9] Standard (VP/mL) |
HEK293 cell lysate (1 mg/ml) |
---|---|---|
Recovery |
||
1 |
1.14E+11 |
94% |
2 |
5.51E+10 |
100% |
3 |
2.60E+10 |
94% |
|
Mean CV |
96% |
Cross reactivities were assessed using other serotypes from the AAV family. Standard curves were generated for each serotype using AAV capsids diluted in the kit diluent.
5 µL of capsids were transferred into a white detection plate (384 low volume), followed by 5 µL of diluent and 10 µL of the HTRF AAV9 capsid detection reagents. The HTRF signal was recorded after an overnight incubation at room temperature.
Signals were interpolated on the assay standard curve to interpolate concentrations. The assay is specific to AAV9 as other serotypes were not detected using this assay.
Empty AAVs Particles |
Cross Reactivity |
---|---|
AAV9 |
100% |
AAV1 |
No |
AAV2 |
No |
AAV3B |
No |
AAV5 |
No |
AAV6 |
No |
AAV8 |
No |
To demonstrate the detection of both full and empty AAV9 capsids, recognition of full AAV9-CMV-eGFP and empty AAV9 capsids were analyzed in the assay. A large range of AAV9-CMV-eGFP concentration (GC/mL) was measured in VP/mL using an independent sample quantitation assay.
5 µL of full or empty capsids diluted in the kit diluent were then transferred into a white detection plate (384 low volume), followed by 5 µL of diluent and 10 µL of the HTRF AAV9 capsid detection reagents. The HTRF signal was recorded after an overnight incubation at room temperature.
As expected, the HTRF AAV9 capsid detection assay can detect in the same way both full and empty AAV9 capsids.
Discover the versatility and precision of Homogeneous Time-Resolved Fluorescence (HTRF) technology. Our HTRF portfolio offers a...