Skip to main content
Menu
US
Application Note

Controlling sample temperature during bead milling using passive cooling with the Omni Bead Ruptor Elite 7 mL tube carriage

passive512

Bead mill sample homogenization, exemplified by the Omni Bead Ruptor Elite™ homogenizer, is a rapid and efficient method for sample disruption. However, for longer processing, the kinetic energy generated can increase sample temperature to potentially affect heat-sensitive analytes. We evaluated the passive cooling effect of the 7 mL carriage on larger volume samples, with the carriage and buffers pre-cooled to various temperatures. The results showed that using a carriage equilibrated at -80 °C in combination with chilled buffers effectively kept processing temperatures below 15 °C for up to 1 minute of processing time for larger volume samples though for samples that are in 2mL tubes, active chamber cooling is preferred utilizing a cryo cooling unit.

For research use only. Not for use in diagnostic procedures.

To view the full content please answer a few questions

By submitting my personal data, I acknowledge that Revvity Inc. and its affiliates (“Company”) will process my personal data provided above consistent with the Company’s Privacy Policy available here.

This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Download Resource

Controlling sample temperature during bead milling using passive cooling with the Omni Bead Ruptor Elite 7 mL tube carriage

Download Application Note